Analog Interface Board

MTR-420MA

JP|EN|CN

外観

Specifications

Name Analog interface board
Model MTR-420MA-PW
External dimensions W74 mm x D63 mm x H58 mm
Output terminal 8-pin screw terminal
Output data  Chamber temperature, Door open / close, Power failure alarm
Output Current Range 4 mA to 20 mA
Maximum load resistance  800 Ω
Allowable contact capacity  DC 30 V, 2A
Power source DC 5 V (supplied from the equipment)
Weight 0.14 kg

Supported Equipment / Output Specifications

機器カテゴリ Models DIP switch
Setting No.
DIP switch bit
Setting (1-4)
Internal Temperature
(Output Range)
Door Open/Close*5 Power Failure Alarm
Temp1 Temp2
超低温フリーザー MDF-DU503VH/703VH series 1 0001 -90℃~38℃
MDF-DU503VXH/703VXH series 1 0001 -90℃~38℃
MDF-DU302VX/502VX/702VX series 1 0001 -90℃~38℃
MDF-DU502VH/702VH series 1 0001 -90℃~38℃
MDF-DU500ZH/700ZH series 1 0001 -90℃~38℃
MDF-DU901VH series 1 0001 -90℃~38℃
MDF-DU300H 3 0011 -90℃~38℃
MDF-U33V/54V/74V series 5 0101 -90℃~38℃
MDF-C8V1 3 0011 -90℃~38℃
MDF-DC200V 3 0011 -90℃~38℃
MDF-DC102VH/202VH series 2 0010 -90℃~38℃
MDF-DC500VX/700VX series 2 0010 -90℃~38℃
バイオメディカルフリーザー MDF-MU300H/500H series*1 8 1000 -50℃~30℃
MDF-MU339/539 series 8 1000 -50℃~30℃
MDF-MU339H/539H series 8 1000 -50℃~30℃
MDF-MU549DH 9 1001 -50℃~30℃
(Upper chamber)
-50℃~30℃
(Lower chamber)
MDF-U443 7 0111 -50℃~30℃
MDF-137/237/437 series 6 0110 -50℃~30℃
薬用保冷庫*4 MPR-S1201XH series 13 1101 -5℃~35℃
MPR-715F 14 1110 -5℃~35℃
(Refrigerator chamber)
-50℃~30℃
(Freezer chamber)
MPR-N250FH series 15 1111 -5℃~35℃
(Refrigerator chamber)
-50℃~30℃
(Freezer chamber)
MPR-N450FH/N650FH series 14 1110 -5℃~35℃
(Refrigerator chamber)
-50℃~30℃
(Freezer chamber)
MPR-514/1014 series*1 12 1100 -5℃~35℃
MPR-S150H/S300H/S500H series 13 1101 -5℃~35℃
MPR-722/1412 series*1 12 1100 -5℃~35℃
MPR-N200H series 13 1101 -5℃~35℃
冷凍機付きインキュベーター MIR-154/254/554 series*2、*3 10 1010 -25℃~75℃
MIR-154S/254S series*1 4 0100 -25℃~75℃
  1. A separately sold battery mounting box (MPR-48B1) is required to install the communication interface product.
  2. A separately sold harness (MTR-HRNS) is required to install the communication interface product.
  3. Equipment for Japan support analog output as standard. Please refer to the catalog and instruction manual for details.
  4. To detect power outage alarms, a separately sold power outage alarm battery (MPR-48B2) or battery mounting box (MPR-48B1) is required.
  5. For equipment with multiple doors, the system detects "open" if any one of the doors is open.